In vitro study on the disinfectability of two split-septum needle-free connection devices using different
Steffen Engelhart1, Martin Exner1, Arne Simon2
1Institute for Hygiene and Public Health, Universitätsklinikum Bonn, Germany.
Abstract:
This in vitro study investigated the external disinfection of two needle-free connection devices (NFC) using Octeniderm(®) (spraying and wiping technique) vs. Descoderm(®) pads (wiping technique). The split-septum membrane of the NFC was contaminated with >10(5) CFU K. pneumoniae or S. epidermidis. The efficacy of the disinfection at 30 sec. exposure time was controlled by taking a swab sample and by flushing the NFC with sterile 0.9% sodium chloride solution. Disinfection with octenidine dihydrochloride 0.1 g, 1-Propanol 30.0 g, and 2-Propanol 45.0 g in 100 g solution was highly effective (CFU reduction ≥4 log) against both microorganisms, whereas the use of 63.1 g 2-Propanol in 100 ml solution led to residual contamination with S. epidermidis. Our investigation underlines that (i) in clinical practice disinfection of NFCs before use is mandatory, and that (ii) details of disinfection technique are of utmost importance regarding their efficacy. Our investigation revealed no significant differences between both split-septum NFC types. Clinical studies are needed to confirm a possible superiority of disinfectants with long-lasting residual antimicrobial activity.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene