Related Experiment Video
Updated: Mar 28, 2026

10:17
20 mJ, 1 ps Yb:YAG Thin-disk Regenerative Amplifier
Published on: July 12, 2017
12.1K
Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database
1Marshfield Clinic. christopher.njeh@natmedphysics.com.
Journal of Applied Clinical Medical Physics
|December 25, 2015
Summary
Enhanced dynamic wedges (EDW) improve radiation therapy dose distribution. This study measured EDW output factors for Varian linacs, finding excellent agreement with existing data and providing new 20 MV data.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Dosimetry
Background:
- Physical wedges are used in radiation therapy to compensate for dose inhomogeneity.
- Enhanced dynamic wedges (EDW) were developed by Varian to address limitations of physical wedges.
- Accurate output factors are crucial for reliable treatment planning and dose delivery.
Purpose of the Study:
- To measure Enhanced Dynamic Wedge (EDW) output factors for 6 MV and 20 MV photon energies on a Varian 2300CD linear accelerator.
- To compile and compare measured EDW output factors with published data for various Varian linear accelerator models.
- To assess the validity and consistency of reference databases for EDW output factors.
Main Methods:
- Experimental measurement of EDW output factors (EDWOF) using a Varian 2300CD at 6 MV and 20 MV.
- Systematic literature review to identify published EDWOF data for different Varian linear accelerator models.
- Statistical comparison of measured data against compiled literature values, including mean differences and coefficients of variation.
Main Results:
- Excellent agreement was observed between measured and published 6 MV EDWOF, with percentage differences from 0.01% to 0.57% (mean 0.03%).
- Published EDWOF showed coefficients of variation ranging from 0.17% to 0.85% (6 MV) and 0.1% to 0.9% (18 MV).
- This study presents the first published EDWOF for 20 MV and a comprehensive compendium for various Varian models.
Conclusions:
- Measured EDWOF values demonstrate high consistency across different Varian linear accelerator models and institutions.
- The findings support the validity of reference databases and the potential for establishing standard datasets for dosimetry.
- This work provides crucial data for future commissioning, beam modeling, and quality assurance in radiation therapy.
More Related Videos
Related Concept Videos
Wedges
2.0K
A wedge is a simple machine that serves various purposes, such as adjusting the elevation of structural or mechanical parts, providing stability for heavy objects, and splitting a body into two parts. This versatile tool can amplify an applied force, making it easier to manipulate large or heavy objects.
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
2.0K
Shear on the Horizontal Face of a Beam Element
612
To understand shear on the flat side of a prismatic beam element, consider the vertical and horizontal shearing forces, and the normal forces, acting on the element. The element's upper (U) and lower (L) sections, which are divided by the beam's neutral axis, are examined. The equilibrium of these forces is determined by applying the equilibrium equation, which helps identify the horizontal shearing force. This force is directly related to the bending moments and the cross-section's...
612