Capillary electrochromatography using monoamine- and triamine-bonded silica nanoparticles as pseudostationary phases
Yuto Takeda1, Yuka Hayashi1, Naonori Utamura1
1Faculty of Pharmaceutical Sciences, Kinki University, 3-4-1, Kowakae, Higashi-Osaka, Osaka 577-8502, Japan.
Abstract:
Monoamine- and triamine-bonded silica nanoparticles were prepared using 3-aminopropyltrimethoxysilane and N(1)-(3-trimethoxysilylpropyl)diethylenetriamine, respectively, and used as pseudostationary phases for capillary electrochromatography. The amine-bonded silica nanoparticles were tightly adsorbed on the inner wall of a capillary and generated fast electro-osmotic flow (2.59 × 10(-4) cm(2) V(-1) s(-1)) toward the anode in an electric field. The electro-osmotic velocities obtained with 20 nm triamine-bonded silica were three to five times larger than those generated by a fused silica capillary and two times faster than those for the commercial cationic polymer-modified capillary. Fast electro-osmotic flow enables rapid analysis. This method was applied to hydrophilic interaction chromatography (HILIC) mode separation of various samples including the size separation of glucose oligomer derivatives and the resolution of four nucleic acid bases.
More Related Videos
07:54Capillary Electrophoresis Mass Spectrometry Approaches for Characterization of the Protein and Metabolite Corona Acquired by Nanomaterials
Published on: October 27, 2020
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
Ion-Exchange Chromatography
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Capillary Electrophoresis: Instrumentation
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...