Actuator pin fracture in magnetically controlled growing rods: two cases
Conor S Jones1, Oliver M Stokes2, Salil B Patel1
1Exeter Spine Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon and Exeter NHS Foundation Trust, Barrack Rd, Exeter EX2 5DW, United Kingdom; Peninsula College of Medicine and Dentistry, Barrack Rd, Exeter EX2 5DW, United Kingdom.
Lengthening pin fractures can cause magnetically controlled growing rod (MCGR) failure in early-onset scoliosis (EOS) patients. Early detection through imaging is crucial for timely intervention and preventing complications.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Pediatric spine deformities
Background:
- Magnetically controlled growing rods (MCGRs) are utilized for non-invasive treatment of early-onset scoliosis (EOS).
- These devices employ a telescopic actuator stimulated by an external remote controller for outpatient lengthening.
- MCGRs offer a less invasive alternative to traditional growing techniques for managing spinal deformities in children.
Observation:
- This report details two cases of MCGR failure attributed to lengthening pin fractures in dual-rod constructs.
- A 12-year-old boy experienced a "popping" sensation and grating, leading to the discovery of a fractured pin.
- Review of other patients revealed bilateral fractures in an asymptomatic girl, indicating a potential subtle failure mode.
Findings:
- The study identified lengthening pin fracture as a subtle cause of MCGR failure in dual-rod systems.
- Radiographic evaluation is essential for diagnosing this specific type of implant failure.
- Bilateral fractures were observed, suggesting a potential systemic issue with the implants or surgical technique.
Implications:
- Clinicians must maintain a high index of suspicion for implant failure in patients with suggestive symptoms or unsuccessful distractions.
- Prompt radiographic evaluation and review of prior imaging are recommended upon suspicion of MCGR failure.
- Regular radiographic follow-up (every 6 months) is advised for centers using ultrasound to monitor MCGR lengthening, ensuring implant structural integrity.
Related Concept Videos
Potential Due to a Magnetized Object
The vector...
Temperature Dependent Deformation
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Plastic Deformation in Circular Shafts
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Force On A Current Loop In A Magnetic Field