Related Experiment Video
Updated: Mar 28, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Cardiac dose reduction during tangential breast irradiation using deep inspiration breath hold: a dose comparison
Ji Hyeon Joo1, Su Ssan Kim2, Seung Do Ahn3
1Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-ro 43-gil, Songpa-gu, Seoul, 138-736, Republic of Korea. hi_juji@daum.net.
Background:
Radiation therapy (RT) for a left-sided breast cancer often involves some incidental exposure of the heart and increase in the rate of major coronary events. One method to reduce the dose to the heart during a tangential breast irradiation is the deep inspiration breath hold (DIBH) technique. Our department adopted DIBH for selected left breast cancer patients with a maximum cardiac distance ≥ 10 mm. We evaluated the effect of the DIBH on cardiac dose compared to normal free breathing (FB). The secondary objective of our present study was to use modeled risk estimates to quantify the risk of coronary events after RT with DIBH.
Methods And Materials:
Thirty-two patients who underwent RT with DIBH at our hospital were retrospectively analyzed. For each patient, two computed tomography (CT) scans were acquired, FB-CT and DIBH-CT. Using a deformable image registration tool, the target volume was deformed from DIBH-CT to FB-CT, and conventional tangential treatment planning was performed, focusing on the equality of target coverage between the two plans. Doses to the heart, left anterior descending (LAD) artery, and ipsilateral lung were assessed.
Results:
By using DIBH, the average mean heart dose was reduced from 724.1 cGy to 279.3 (p < 0.001). The relative heart volume irradiated with 10 Gy-50 Gy was consistently reduced. The mean dose to the LAD coronary artery was reduced from 4079.1 cGy to 2368.9 cGy (p < 0.001). The ipsilateral lung volume receiving 20 Gy or more and 40 Gy or more was reduced by 2.2 % in both cases. Estimated risks of coronary events at 10 years were 4.03 and 2.55 % for RT with FB and DIBH, respectively (p < 0.001).
Conclusions:
The use of DIBH during RT of the left-sided breast considerably reduces the doses delivered to the heart and LAD artery with similar target coverage. For the current study patients, the probability of major coronary events was reduced with DIBH.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Cardiopulmonary Resuscitation II: ACLS Airway Management