A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
K2-EDTA and K3-EDTA Greiner Tubes for HbA1c Measurement.
Alen Vrtaric1, Petra Filipi2, Marina Hemar2
1University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia alenvrtaric@gmail.com.
K2-ethylenediaminetetraacetic acid (EDTA) and K3-EDTA Greiner tubes are interchangeable for glycosylated hemoglobin (HbA1c) testing. This study found no significant clinical or statistical differences, ensuring reliable HbA1c measurements with either tube type.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Hematology
Background:
- Glycosylated hemoglobin (HbA1c) is a critical biomarker for diabetes mellitus management.
- Accurate HbA1c measurement relies on appropriate sample collection and anticoagulant choice.
- Different forms of EDTA anticoagulants (K2-EDTA vs. K3-EDTA) are used in blood collection tubes.
Purpose of the Study:
- To evaluate the interchangeability of K2-EDTA and K3-EDTA Greiner tubes for HbA1c analysis.
- To determine if K2-EDTA and K3-EDTA tubes yield comparable results using the Abbott ARCHITECT HbA1c assay.
- To assess the clinical and statistical significance of any observed differences in HbA1c measurements between the two tube types.
Main Methods:
- Blood samples were collected from 45 outpatients into both K2-EDTA and K3-EDTA Greiner Vacuette tubes.
- HbA1c concentrations were measured using the Abbott ARCHITECT chemiluminescent microparticle assay on the ARCHITECT i2000SR immunoanalyzer.
- Statistical analysis included Wilcoxon test and Passing-Bablok regression to compare results from the two tube types.
Main Results:
- Median HbA1c levels were 54 mmol/mol for K2-EDTA and 56 mmol/mol for K3-EDTA tubes.
- No statistically significant difference was observed between K2-EDTA and K3-EDTA tubes (bias = -1.29 mmol/mol, P = 0.24).
- Passing-Bablok regression analysis indicated no constant or proportional error, with a slope of 1.00 (95% CI[0.98 to 1.06]).
Conclusions:
- K2-EDTA and K3-EDTA Greiner tubes demonstrate no clinically significant bias for HbA1c measurement.
- The findings support the safe and interchangeable use of both K2-EDTA and K3-EDTA tubes for HbA1c determination with the Abbott ARCHITECT assay.
- Laboratory protocols can confidently utilize either K2-EDTA or K3-EDTA tubes for routine HbA1c testing.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Auxiliary Complexing Reagents
Complexometric EDTA Titration Curves
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...