A Case of Useful Short-Spaced Bipolar Pacing of a Left Ventricular Lead to Avoid Phrenic Nerve Stimulation
Tomoyuki Kabutoya1, Yasushi Imai, Hiroaki Watanabe
1Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine.
Abstract:
A 48-year-old woman underwent cardiac resynchronization therapy defibrillator implantation. Coronary sinus (CS) venography showed only one adequate anterior branch for a left ventricular lead. We were able to introduce a quadripolar left ventricular lead (Medtronic 4398-88 cm) to the distal portion of the anterior branch. Although phrenic nerve stimulation (PNS) occurred due to distal bipolar pacing (distal 1-mid 2, with 21-mm distance) and proximal pacing (mid 3-proximal 4, distance 21mm), short-spaced bipolar pacing (mid 2-3, distance 1.3 mm) did not induce PNS until 9V pacing. Shared bipolar pacing from each left ventricular electrode (distal 1 to proximal 4) as cathode and a right ventricular (RV) coil as anode resulted in PNS by 3.0V at 0.4 ms. Although quadripolar pacing could avoid PNS by switching the pacing site (ie, from distal bipolar to proximal bipolar), it might not avoid PNS in cases where the phrenic nerve and CS branch are parallel and in close proximity. We found that even though the phrenic nerve and CS branch were parallel and close, short-spaced bipolar pacing could avoid PNS. In conclusion, short-spaced bipolar pacing selected by quadripolar pacing might be beneficial to avoid PNS when the implantable branch is limited.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...