Hospital steam sterilizer usage: could we switch off to save electricity and water?
Forbes McGain1, Graham Moore2, Jim Black3
1Departments of Anaesthesia and Intensive Care, Western Health, Footscray Hospital, Australia forbes.mcgain@wh.org.au.
Objectives:
Steam sterilization in hospitals is an energy and water intensive process. Our aim was to identify opportunities to improve electricity and water use. The objectives were to find: the time sterilizers spent active, idle and off; the variability in sterilizer use with the time of day and day of the week; and opportunities to switch off sterilizers instead of idling when no loads were waiting, and the resultant electricity and water savings.
Methods:
Analyses of routine data for one year of the activity of the four steam sterilizers in one hospital in Melbourne, Australia. We examined active sterilizer cycles, routine sterilizer switch-offs, and when sterilizers were active, idle and off. Several switch-off strategies were examined to identify electricity and water savings: switch off idle sterilizers when no loads are waiting and switch off one sterilizer after 10:00 h and a second sterilizer after midnight on all days.
Results:
Sterilizers were active for 13,430 (38%) sterilizer-hours, off for 4822 (14%) sterilizer-hours, and idle for 16,788 (48%) sterilizer-hours. All four sterilizers were simultaneously active 9% of the time, and two or more sterilizers were idle for 69% of the time. A sterilizer was idle for two hours or less 13% of the time and idle for more than 2 h 87% of the time. A strategy to switch off idle sterilizers would reduce electricity use by 66 MWh and water use by 1004 kl per year, saving 26% electricity use and 13% of water use, resulting in financial savings of AUD$13,867 (UK£6,517) and a reduction in 79 tonnes of CO2 emissions per year. An alternative switch-off strategy of one sterilizer from 10:00 h onwards and a second from midnight would have saved 30 MWh and 456 kl of water.
Conclusions:
The methodology used of how hospital sterilizer use could be improved could be applied to all hospitals and more broadly to other equipment used in hospitals.
More Related Videos
06:21Thermal Behavior and Power Efficiency Comparison of AC vs. DC Electrical Heating in a Distillation Column Using Infrared Thermography Analysis
Published on: December 5, 2025
08:41Microalgae Cultivation and Biomass Quantification in a Bench-Scale Photobioreactor with Corrosive Flue Gases
Published on: December 19, 2019
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Bioreactor Controls-II
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
The Carnot Cycle
What could be the theoretical limit to the efficiency of a heat engine? The...