Related Experiment Video
Updated: Mar 26, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Reliability of patient specific instrumentation in total knee arthroplasty
Abstract:
The aim of this study was to compare the precision between Patient Specific Instrumentation (PSI) and Conventional Instrumentation (CI) as determined intra-operatively by a pinless navigation system. Eighty patients were included in this prospective comparative study and they were divided into two homogeneous groups. We defined an original score from 6 to 30 points to evaluate the accuracy of the position of the cutting guides. This score is based on 6 objective criteria. The analysis indicated that PSI was not superior to conventional instrumentation in the overall score (p = 0.949). Moreover, no statistically significant difference was observed for any individual criteria of our score. Level of evidence II.

