Comparison of dosimetry methods for panoramic radiography: thermoluminescent dosimeter measurement versus personal
Chena Lee1, Sam-Sun Lee1, Jo-Eun Kim1
1Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.
Objective:
The purpose of this study was to evaluate the patient radiation dose based on panoramic radiography and calculated with personal computer-based Monte Carlo (PCXMC) software compared with thermoluminescent dosimetry (TLD) measurement. We also proposed appropriate input values for dose-determining factors in PCXMC.
Study Design:
Tissue-absorbed doses and the effective dose based on panoramic radiography were measured with TLD and with PCXMC under various conditions. The calculated PCXMC doses were compared with those measured with TLD.
Results:
The effective doses calculated with PCXMC were higher by 9.55% to 51.24% compared with the doses measured with TLD. Reference points on the Y-axis and Z-axis were the sensitive factors when calculating the effective dose. The differences between the highest and the lowest organ doses were 0.32 and 0.10 mGy, respectively, for PCXMC calculation and TLD measurement.
Conclusions:
The effective dose calculated with PCXMC was generally higher than the dose measured by using TLD, and the absorbed doses varied by organ more severely in the PCXMC calculations than in the TLD measurements. The effective dose obtained from PCXMC calculations was dependent on input values for dose-determining factors. Standard values for each dose-determining factor required for the application of PCXMC to panoramic radiography were suggested in this study.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation