Setting up of in-vacuum PIXE system for direct elemental analysis of thick solid environmental samples
M S Rihawy1, I M Ismail1, D Halloum1
1Department of Chemistry, Atomic Energy Commission, P.O. Box 6091, Damascus, Syrian Arab Republic.
Abstract:
Experimental set-up, development, characterization, and calibration of an in-vacuum PIXE system at the tandem accelerator facility of the Atomic Energy Commission of Syria (AECS) is described. The PIXE system calibration involved experimental characterization of the X-ray detector parameters and careful determination of the H-values that control dependence of the detector solid angle with the X-ray energies and correct imperfect values of the detector efficiency. Setting up of an electron flood gun to compensate charge built up and utilization of a beam profile monitor to perform indirect measurement of the beam charge, provide a direct PIXE measurement of thick insulating samples in-vacuum. The PIXE system has been subsequently examined to verify its ability to perform direct PIXE measurements on geological materials. A combination of minimum sample preparation procedures and specific experimental conditions applied enables simple and accurate elemental analysis. Elemental concentrations of several elements heavier than sodium in different reference geological samples, at about 5-10% absolute accuracy for most elements, have been determined. Comprehensive discussion of the obtained elemental concentration values, for most elements of visible X-ray peaks in the PIXE spectra, has been considered.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Lab
Preparation of Samples for Electron Microscopy
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Instrumentation