The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Cricoid pressure impedes tracheal intubation with the Pentax-AWS Airwayscope®: a prospective randomized trial
N Komasawa1, H Kido2, Y Miyazaki2
1Department of Anaesthesiology, Osaka Medical College, Daigaku-machi 2-7, Takatsuki, Osaka 569-8686, Japan ane078@poh.osaka-med.ac.jp.
Background:
It is unclear how cricoid pressure affects tracheal intubation with the Pentax-AWS Airwayscope(®) (AWS). We conducted a prospective randomized clinical trial in anaesthetized patients.
Methods:
Sixty patients were allocated to either the cricoid pressure (CP) group (n=30) or the sham group (n=30). We compared the two groups with regard to intubation time, number of attempts required for insertion of the Intlock blade (disposable blade of the AWS) and tracheal intubation, percentage of glottic opening (POGO) score, and subjective difficulty of both laryngoscopy and passage of a tube through the glottis.
Results:
Intubation time was significantly longer in the CP group (median 45[IQR40-59] s) than in the sham group (32[28-45] s) (P=0.003, 95% CI for median difference 5-24 s). The number required for insertion of the Intlock blade did not differ between the groups (P=0.08), but the number for tracheal intubation was significantly higher in the CP group (1 attempt in 14 patients, 2 in 7, 3 in 9) than in the sham group (1 attempt in 24 patients, 2 in 6; P=0.002). POGO score did not differ significantly between the groups (P=0.60), nor did the subjective difficulty of laryngoscopy (P=0.06). The visual analogue scale score for passage of a tube through the glottis was significantly higher in the CP group than in the sham group (P<0.001).
Conclusions:
Cricoid pressure impedes tracheal intubation using the AWS, and is associated with longer intubation time, which can be attributed to increased difficulty in the passage of a tube through the glottis.
Clinical Trial Registry Number:
UMIN000018209.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques