Related Experiment Video
Updated: Mar 26, 2026

A Model of Acute Lung Injury Following Visceral Ischemia-Reperfusion by Supra-Coeliac Aortic Cross Clamping in Rats
Published on: August 15, 2025
Suprarenal fixation resulting in intestinal malperfusion after endovascular aortic aneurysm repair
Andrea Siani1, Federico Accrocca2, Gennaro De Vivo2
1Vascular and Endovascular Surgery Unit, San Paolo Hospital, Rome, Italy lucamaria.siani@tin.it.
Abstract:
Superior mesenteric artery (SMA) and coeliac axis (CA) occlusion after endovascular abdominal aneurysm aortic repair (EVAR-AAA), using endograft with suprarenal fixation, are uncommon. However, we are reporting a case of visceral malperfusion, which occurred 7 days after successful EVAR with suprarenal fixation for symptomatic AAA. Endograft metal stent barbs caused severe stenosis of SMA and CA. A successful recovery of SMA was carried out by means of a balloon-expandable stent released through bare metal stent barbs. We believe that an unfavourable anatomy of a proximal aortic neck and visceral aorta may have caused a wrong stent strut deployment with the coverage of CA and SMA.
Related Concept Videos
Aneurysm III: Interprofessional Care
Aneurysm IV: Nursing Management
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Aneurysm I: Introduction