Evaluating the Effects of Different Polishing Methods on Color Stability of Dental Restorations in Pediatric Dentistry
Published on: June 6, 2025
586
Dental color matching: A comparison between visual and instrumental methods
Christopher Igiel1, Michael Weyhrauch, Stefan Wentaschek
1Department of Prosthetic Dentistry, University Medical Center.
Dental Materials Journal
|February 3, 2016
Summary
This study evaluated three dental color-measuring devices against visual shade matching. Results showed that while devices like Shadepilot (SP) offered some clinical acceptability, color differences often exceeded acceptable thresholds for dental restorations.
Area of Science:
- * Dental Materials Science
- * Restorative Dentistry
- * Colorimetry
Background:
- * Accurate dental shade selection is crucial for aesthetic outcomes in restorative dentistry.
- * Visual shade matching is subjective and prone to operator variability.
- * Instrumental color measurement aims to improve accuracy and reproducibility.
Purpose of the Study:
- * To compare the agreement rate and color difference (ΔE*ab) of three dental color-measuring devices (Shadepilot, CrystalEye, ShadeVision) with visual shade identification.
- * To assess the clinical acceptability of color differences obtained through instrumental and visual methods.
Main Methods:
- * Three dental color-measuring devices (Shadepilot (SP), CrystalEye (CE), ShadeVision (SV)) were used to measure tooth color.
- * Two operators performed visual shade selection using VITA classic shade tabs.
- * Statistical analyses included agreement rate (%), color difference (ΔE*ab), McNemar test, Student's t-test, and Bland Altman scatterplots.
Main Results:
- * Agreement rates with visual shade determination were: SP 56.3%, CE 49.0%, SV 51.3%.
- * Color differences (ΔE*ab) between visually and instrumentally selected shades and natural teeth frequently exceeded the threshold for acceptability.
- * For Shadepilot, ΔE*ab values fell within a range of clinical acceptability when comparing both methods.
Conclusions:
- * Current dental color-measuring devices show variable agreement with visual shade matching.
- * Achieving clinically acceptable color matches remains a challenge, with significant color differences often observed.
- * Further advancements in dental colorimetry are needed to consistently meet aesthetic demands.