Related Experiment Video
Updated: Mar 26, 2026

Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
Time-related changes in neointimal tissue coverage of a novel Sirolimus eluting stent: Serial observations with
Gioel Gabrio Secco1, Alessio Mattesini2, Rossella Fattori3
1Interventional Cardiology, "Santi Antonio e Biagio e Cesare Arrigo" Hospital, Alessandria, Italy; NIHR Biomedical Research Unit, Royal Brompton & Harefield NHS Foundation Trust, London, UK.
Background:
DES has reduced rates of restenosis compared with BMS but it has been associated with delayed healing and increase of stent thrombosis. The aim of our study was to evaluate the vascular time-related changes following implantation of a new SES coated with an amorphous silicon carbide that allows faster re-endothelisation (Orsiro-Biotronik).
Methods:
This prospective registry enrolled STEMI-patients with multi-vessel disease, thus candidates for a two-step procedure. PCI of the culprit lesion was performed with at least one Orsiro stent that was OCT-analysed during the second-step procedure (deferred to 30,90 and 180-days).
Results:
16 of the 95 patients with MVD underwent an OCT evaluation of the study device implanted in the culprit lesion during the second-step staged procedure and were enrolled in the present Registry. A total of 3060-struts were analysed. The percentage of uncovered struts was 19.6% at30-days, 1.3% at90-days and 1.8% at180-days (p<0.001). The percentage of cross section with ≥1 uncovered struts were 51.3% at30-days, 6.5% at90-days and 5.7% at180-days (p<0.001). The percentage of cross sections containing thrombus was 6.2% at30-days while no thrombus was detected both at90 and 180-days.
Conclusions:
Our data shows that the new Orsiro stent promotes early and persistent strut coverage with low peri-strut thrombus. This pilot OCT evaluation might suggest a low incidence of late adverse events and anticipate safe outcome after early withdrawal of dual antiplatelet therapy.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
11:45Generation and 3-Dimensional Quantitation of Arterial Lesions in Mice Using Optical Projection Tomography
Published on: May 26, 2015