Related Experiment Video
Updated: Mar 26, 2026

Use of Enzymatic Biosensors to Quantify Endogenous ATP or H2O2 in the Kidney
Published on: October 12, 2015
A novel conductometric biosensor based on hexokinase for determination of adenosine triphosphate
I S Kucherenko1, D Yu Kucherenko2, O O Soldatkin2
1Institute of Molecular Biology and Genetics of National Academy of Sciences of Ukraine, 150 Zabolotny Str., Kyiv 03680, Ukraine; Taras Shevchenko National University of Kyiv, 64 Volodymyrska str., 01003 Kyiv, Ukraine; Institute of Analytical Sciences, UMR5280 CNRS/UCBL/ENS, 5 rue de la Doua, 69100, Villeurbanne, France.
Abstract:
The paper presents a simple and inexpensive reusable biosensor for determination of the concentration of adenosine-5'-triphosphate (ATP) in aqueous samples. The biosensor is based on a conductometric transducer which contains two pairs of gold interdigitated electrodes. An enzyme hexokinase was immobilized onto one pair of electrodes, and bovine serum albumin-onto another pair (thus, a differential mode of measurement was used). Conditions of hexokinase immobilization on the transducer by cross-linking via glutaraldehyde were optimized. Influence of experimental conditions (concentration of magnesium ions, ionic strength and concentration of the working buffer) on the biosensor work was studied. The reproducibility of biosensor responses and operational stability of the biosensor were checked during one week. Dry storage at -18 °C was shown to be the best conditions to store the biosensor. The biosensor was successfully applied for measurements of ATP concentration in pharmaceutical samples. The proposed biosensor may be used in future for determination of ATP and/or glucose in water samples.