Long-term biventricular HeartWare ventricular assist device support--Case series of right atrial and right
Sajad Shehab1, Peter S Macdonald2, Anne M Keogh2
1Centre for Cardiovascular and Chronic Care, Faculty of Health, University of Technology Sydney, Ultimo; Heart Transplant Unit, St. Vincent's Hospital, Darlinghurst.
Background:
There is limited information on outcomes using the HeartWare ventricular assist device (HVAD; HeartWare, Framington, MA) as a biventricular assist device, especially with respect to site of right ventricular assist device (RVAD) implantation.
Methods:
Outcomes in 13 patients with dilated cardiomyopathy and severe biventricular failure who underwent dual HVAD implantation as bridge to transplantation between August 2011 and October 2014 were reviewed.
Results:
Of 13 patients, 10 were Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) Level 1, and 3 were INTERMACS Level 2. Mean age was 45 ± 11 years, and mean body mass index was 26 ± 4 kg/m(2). There were 7 patients on temporary mechanical support pre-operatively (extracorporeal life support, n = 5; intra-aortic balloon pump, n = 2). The median hospital length of stay was 53 days (interquartile range [IQR] 33-70 days) with a median intensive care unit length of stay of 14 days (IQR 8-36 days). The median length of support on device was 269 days (IQR 93-426 days). The right HVAD was implanted in the right ventricular (RV) free wall in 6 patients and in the right atrial (RA) free wall in 7 patients. Transplantation was successfully performed in 5 patients, and overall survival for the entire cohort was 54%. RVAD pump thrombosis occurred in 3 of 6 RV pumps and 1 of 7 RA pumps. No left ventricular assist device pump thrombosis was observed. Bleeding tended to be higher in the RV implantation group (3 of 6 vs 0 of 7). During follow up, 6 patients died (4 of 7 in the RA group vs 2 of 6 in the RV group). Cause of death was multiple-organ failure in 3 patients, sepsis in 2 patients, and intracerebral hemorrhage in 1 patient.
Conclusions:
Critically ill patients who require biventricular support can be successfully bridged to transplant using 2 HVADs. RA implantation may allow right heart support with lower pump thrombosis and bleeding complications, although this was at the expense of a higher mortality in this cohort.
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Heart Failure VI: Adjunct Therapies
Heart Failure V: Medical Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mitral Stenosis III: Medical Management