Open-Source Real-Time Closed-Loop Electrical Threshold Tracking for Translational Pain Research
Published on: April 21, 2023
Threshold Evolution as an Analysis of the Different Pulse Frequencies in Rechargeable Systems for Spinal Cord
David Abejón1, Pablo Rueda2, Ricardo Vallejo3,4,5
1Head of Pain Unit, Hospital Universitario Quirón Madrid, Madrid, Spain.
Introduction:
Pulse frequency (Fc) is one of the most important parameters in neurostimulation, with Pulse Amplitude (Pw) and Amplitude (I). Up to certain Fc, increasing the number of pulses will generate action potentials in neighboring neural structures and may facilitate deeper penetration of the electromagnetic fields. In addition, changes in frequency modify the patient's sensation with stimulation.
Materials/Methods:
Fifty patients previously implanted with rechargeable current control spinal cord stimulation. With pulse width fixed at 300 μsec, we stimulated at 26 different Fc values between 40 and 1200 Hz and determine the influence of these changes on different stimulation thresholds: perception threshold (Tp ), therapeutic perception (Tt), and discomfort threshold (Td). Simultaneously, paresthesia coverage of the painful area and patient's sensation and satisfaction related to the quality of stimulation were recorded.
Results:
Pulse Fc is inversely proportional to stimulation thresholds and this influence is statistically significant (p < 0.05). As Pulse Fc increased from 40 to 1200 Hz, the mean threshold decreases from 7.25 to 1.38 mA (Tp ), 8.17 to 1.63 (Tt ), and 9.20 to 1.85 (Td). Significant differences for Tp and Tt began at 750 Hz (Tp , Tt ) and at 650 Hz for Td. No significant influence was found regarding paresthesia coverage. As expected, Fc affects significantly patient's sensation and satisfaction.
Discussion:
Changes in Fc affect the quality of paresthesias. Within the evaluated parameters higher frequencies are inversely proportional to stimulation thresholds and Tt. It seems that Fc is a vital parameter to achieve therapeutic success.
Conclusions:
Changes in Fc is a useful parameter to modulate the patient's sensory perception. Fc can be successfully used to adjust the quality of the paresthesias and to modify patient's subjective sensation. We showed that as the frequency increases, the patient's satisfaction with the perceived sensation decreases, suggesting that higher Fc may need to be set up at subthreshold amplitude to achieve positive response.