A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Half-time bone scintigraphy in prostate and breast cancer patients
Willem Grootjans1, Sara J Serém2, Monique I Gomes2
1Department of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands - w.grootjans@lumc.nl.
Background:
Developments in image reconstruction techniques for planar imaging, also known as enhanced planar processing (EPP), enable the possibility to reconstruct planar scintigraphic images with low count statistics, providing the opportunity to reduce image acquisition time. In this study, the performance of EPP for oncologic half-time bone scintigraphy images was evaluated.
Methods:
The EPP software was evaluated for different imaging conditions using standardized phantom experiments. Additionally, 51 patients with prostate and breast cancer were prospectively included and underwent bone scintigraphy using a standard and half-time protocol. Independent reading was performed on three image types (standard, half-time non-processed, and half-time EPP) by three observers, scoring the number and anatomical location of lesions, image quality, and diagnostic confidence by which the definitive diagnosis was made.
Results:
EPP images had improved contrast and lower noise levels compared to the non-processed half-time images. It was determined that EPP images acquired at double scan speed had similar image quality to the standard non-processed images. There was substantial agreement with respect to diagnosis and diagnostic confidence based on all three image types between the observers. Image quality in the EPP images was higher with respect to the non-processed half-time images, and was comparable to the standard images.
Conclusions:
Diagnostic confidence was not affected by reduction in image acquisition time. There was substantial agreement between all three observers with respect to the diagnosis provided in all three image types. Subjective and objective image quality improved when half-time images were processed with EPP software.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...