Related Experiment Video
Updated: Mar 25, 2026

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Feasibility and Acute Hemodynamic Effect of Left Ventricular Septal Pacing by Transvenous Approach Through the
Masih Mafi-Rad1, Justin G L M Luermans1, Yuri Blaauw1
1From the Department of Cardiology, Maastricht University Medical Center, The Netherlands (M.M.-R., J.G.L.M.L., Y.B., M.J., H.J.C., K.V.); and Department of Physiology, Maastricht University, Cardiovascular Research Institute Maastricht, The Netherlands (F.W.P.).
Background:
Left ventricular septal (LVS) pacing reduces ventricular dyssynchrony and improves cardiac function relative to right ventricular apex (RVA) pacing in animals. We aimed to establish permanent placement of an LVS pacing lead in patients using a transvenous approach through the interventricular septum.
Methods And Results:
Ten patients with sinus node dysfunction scheduled for dual-chamber pacemaker implantation were prospectively enrolled. A custom pacing lead with extended helix was introduced via the left subclavian vein and, after positioning against the right ventricular septum (RVS) using a preshaped guiding catheter, driven through the interventricular septum to the LVS. The acute hemodynamic effect of RVA, RVS, and LVS pacing was evaluated by invasive LVdP/dtmax measurements. The lead was successfully delivered to the LVS in all patients. Procedure time and fluoroscopy time shortened with experience. QRS duration was shorter during LVS pacing (144 ± 20 ms) than during RVA (172 ± 33 ms; P = 0.02 versus LVS) and RVS pacing (165 ± 17 ms; P = 0.004 versus LVS). RVA and RVS pacing reduced LVdP/dtmax compared with baseline atrial pacing (-7.1 ± 4.1% and -6.9 ± 4.3%, respectively), whereas LVS pacing maintained LVdP/dtmax at baseline level (1.0 ± 4.3%; P = 0.001 versus RVA and RVS). R-wave amplitude and pacing threshold were 12.2 ± 6.7 mV and 0.5 ± 0.2 V at implant and remained stable during 6-month follow-up without lead-related complications.
Conclusions:
Permanent placement of an LVS pacing lead by transvenous approach through the interventricular septum is feasible in patients. LVS pacing preserves acute left ventricular pump function. This new pacing method could serve as an alternative and hemodynamically preferable approach for antibradycardia pacing.