Related Experiment Video
Updated: Mar 25, 2026

10:12
Target Cell Pre-enrichment and Whole Genome Amplification for Single Cell Downstream Characterization
Published on: May 15, 2018
9.6K
Evaluation of the Check-Points Check MDR CT103 and CT103 XL Microarray Kits by Use of Preparatory Rapid Cell Lysis
Scott A Cunningham1, Shawn Vasoo2, Robin Patel3
1Division of Clinical Microbiology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota, USA.
Journal of Clinical Microbiology
|February 19, 2016
Abstract:
Using a rapid bacterial lysis method, the Check MDR CT103 and CT103 XL microarrays demonstrated accuracies of 98.1% and 94.2%, respectively, for detection of known resistance genes in 108 multidrug-resistant Gram-negative bacilli. In 45 isolates, 49 previously unrecognized extended-spectrum β-lactamase or plasmid AmpC targets were detected and confirmed by conventional PCR.