Estimation of Radiation Dose in CT Based on Projection Data
Xiaoyu Tian1, Zhye Yin2, Bruno De Man2
1Department of Biomedical Engineering, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, 2424 Erwin Road, Suite 302, Durham, NC, 27705, USA.
Abstract:
Managing and optimizing radiation dose has become a core problem for the CT community. As a fundamental step for dose optimization, accurate and computationally efficient dose estimates are crucial. The purpose of this study was to devise a computationally efficient projection-based dose metric. The absorbed energy and object mass were individually modeled using the projection data. The absorbed energy was estimated using the difference between intensity of the primary photon and the exit photon. The mass was estimated using the volume under the attenuation profile. The feasibility of the approach was evaluated across phantoms with a broad size range, various kVp settings, and two bowtie filters, using a simulation tool, the Computer Assisted Tomography SIMulator (CATSIM) software. The accuracy of projection-based dose estimation was validated against Monte Carlo (MC) simulations. The relationship between projection-based dose metric and MC dose estimate was evaluated using regression models. The projection-based dose metric showed a strong correlation with Monte Carlo dose estimates (R (2) > 0.94). The prediction errors for the projection-based dose metric were all below 15 %. This study demonstrated the feasibility of computationally efficient dose estimation requiring only the projection data.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
