Hypoglycemia Prevention and User Acceptance of an Insulin Pump System with Predictive Low Glucose Management
Pratik Choudhary1, Birthe S Olsen2, Ignacio Conget3
11 King's College London , London, United Kingdom .
Background:
The MiniMed 640G sensor-augmented insulin pump system (Medtronic, Inc., Northridge, CA) can automatically suspend insulin delivery in advance of predicted hypoglycemia and restart it upon recovery. The aims of this analysis were to determine the rate at which predicted hypoglycemia was avoided with this strategy, as well as to assess user acceptance of the system and its insulin management features.
Subjects And Methods:
Forty subjects with type 1 diabetes used the system for 4 weeks. We retrospectively evaluated performance of the system, using downloaded pump and sensor data, and evaluated user acceptance via questionnaires.
Results:
There were 2,322 suspend before low events (2.1 per subject-day). The mean (± SD) duration of pump suspension events was 56.4 ± 9.6 min, and the mean subsequent sensor glucose (SG) nadir was 71.8 ± 5.2 mg/dL. SG values following 1,930 (83.1%) of the predictive suspensions did not reach the preset low limit. Nadir SG values of ≤50 and ≤60 mg/dL were seen in 207 (8.9%) and 356 (15.3%) of the predictive suspensions, respectively. Blood glucose (BG) and SG values before and during the study were comparable (P > 0.05). The mean absolute relative difference between paired SG and BG values was 10.9 ± 13.8%. Subjects felt confident using the system, agreed that it helped protect them from hypoglycemia, and wished to continue using it.
Conclusions:
Automatic insulin pump suspension as implemented in the MiniMed 640G system can help patients avoid hypoglycemia, without significantly increasing hyperglycemia.
More Related Videos
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024
Related Concept Videos
Hypoglycemia and Glucagon
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Oral Hypoglycemic Agents: Glinides
Oral Hypoglycemic Agents: Biguanides and Glitazones
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Type 2 and Gestational