The Effects of Temperature and Relative Humidity on Point-of-Care Glucose Measurements in Hospital Practice in a
Busadee Pratumvinit1, Nattakom Charoenkoop2, Soamsiri Niwattisaiwong2
1Department of Clinical Pathology, Faculty of Medicine Siriraj hospital, Mahidol University, Bangkok, Thailand busadee.pra@mahidol.ac.th.
Background:
Hospitals in tropical countries experience conditions that exceed manufacturer temperature and humidity limits for point-of-care (POC) glucose reagents. Our goal was to assess the effects of out-of-limits storage temperature, operating temperature, and operating humidity on POC glucose measurement reliability.
Methods:
Quality control measurements were performed monthly using glucose test strips stored under controlled conditions and in inpatient wards under ambient conditions. Glucose test strips were evaluated in groups organized by operating temperatures of 24-25 (group 1), 28-29 (group 2), and 33-34°C (group 3), and relative humidity (RH) of ≤70 (group A), ~80 (group B), and ~90% (group C).
Results:
Glucose results for different storage conditions were inconsistent. Measurements at higher operating temperatures had lower values with mean differences of -2.4 (P < .001) and -36.5 (P < .001) mg/dL (28-29 vs 24-25°C), and -3.6 (P < .001) and -37.4 (P < .001) mg/dL (33-34 vs 24-25°C) for low and high control levels, respectively. Measurements at higher RH had lower values with mean differences of -4.0 (P < .001) and -13.2 (P < .001) mg/dL (~80 vs ≤70% RH), and -5.8 (P < .001) and -16.6 (P < .001) mg/dL (~90 vs ≤70% RH) for low and high levels, respectively.
Conclusions:
High temperature and high RH decreased glucose concentrations for the POC oxidase-based system we evaluated. We recommend that individual hospitals perform stress testing, then determine if maximum absolute differences, which represent highest risk for patients, are clinically significant for decision making by using error grid analysis.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Factors Affecting Body Temperature
Factors may include:
Errors occurring during blood pressure monitoring
Several factors...