Limit for qualitative detection of benzoylecgonine in urine using EMIT reagents and the Hitachi 705 automatic
Abstract:
The limit of detection for the cocaine metabolite (hydrolysis product), benzoylecgonine, using Syva EMIT reagents and the Hitachi 705 automatic analyzer has been found to be 10 ng/mL at the 99% confidence level. For this determination 30 controls were prepared from a single urine pool known to be negative for cocaine metabolite. Urine samples of 90 patients were found to be drug-free when analyzed by selected ion monitoring gas chromatography/mass spectrometry for ecgonine methyl ester and benzoylecgonine at sensitivities of less than 0.4 ng/mL and 1.3 ng/mL, respectively. To test matrix effects on the limit of detection, these samples were screened with the Hitachi 705. Results for eight of the known negative samples fell above the 10-ng/mL limit of detection while only one presumptive positive resulted at an arbitrary 20-ng/mL cutoff. Detection limits were also determined using 20 different negative patients' urine samples fortified with benzoylecgonine. At the 99% confidence level using EMIT d.a.u. and 705 reagents the detection limits were found to be 44 ng/mL and 35 ng/mL, respectively. It was found that 62% of patients' samples positive for benzoylecgonine fell below the 300-ng/mL cutoff and above a cutoff at 50 ng/mL.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Qualitative Analysis
There are two main approaches to qualitative analysis:...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors