Comparison of Temperature Programmable Split/Splitless and Cool On-Column Inlets for the Determination of Glycerol
Matthew Giardina1, James D McCurry2
1Gas Phase Separations Division, Agilent Technologies, Inc., 2850 Centerville Road, Wilmington, DE 19080, USA matthew_giardina@agilent.com.
Abstract:
The European Standard EN 14105:2011-07 is an analysis method for quantifying free glycerol and residual mono-, di- and triacylglycerides impurities in biodiesel by gas chromatography. The method specifies an "on-column injector or equivalent device" as the means of sample introduction. Cool on-column (COC) would appear to be an ideal choice, particularly for quantifying triacylglycerides, as it provides high quantitative accuracy and precision with minimal mass discrimination. However, there are a few drawbacks in using COC for this application. The relatively high concentration of the biodiesel in the prepared samples impedes solvent focusing of early eluting compounds such as glycerol, causing band broadening and shifts in retention time compared with the external calibration standards. More problematic is method robustness when using a metal retention gap. Repeated injections onto the retention gap cause the method control specification to fail within relatively few injections. As an alternative, a temperature programmable split/splitless (TPSS) inlet was investigated for performance equivalency. The results demonstrate that the TPSS yields concentration measurements indistinguishable from the COC inlet at the 95% confidence level. In addition, the robustness of the TPSS far exceeds that of the COC inlet by eliminating the performance control failure and providing solvent focusing for the early eluting peaks.
More Related Videos
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
17:39A Lipid Extraction and Analysis Method for Characterizing Soil Microbes in Experiments with Many Samples
Published on: July 16, 2017
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II