Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

European Multicentre Study on Recurrent Imaging Practice in Computed Tomography.

The British journal of radiology·2026
Same author

Improved coronary CT angiography image quality using photon-counting detector CT in SCAPIS reexamination: scan protocol and comparative analysis.

European journal of radiology·2026
Same author

Comparison of photon-counting CT with energy-integrating CT in temporal bone imaging: an anthropomorphic phantom study.

Acta radiologica (Stockholm, Sweden : 1987)·2026
Same author

A methodology for assessing the effect of localizer orientation on the consistency of tube current modulation in CT.

Radiation protection dosimetry·2026
Same author

Feasibility of micro-computed tomography for evaluating anatomical landmarks in temporal bone specimens.

Radiation protection dosimetry·2026
Same author

Annotation and characterization of lesions in breast tomosynthesis images.

Radiation protection dosimetry·2026

Related Experiment Video

Updated: Mar 25, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
06:33

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement

Published on: July 29, 2013

11.8K

Improvements to image quality using hybrid and model-based iterative reconstructions: a phantom study.

Marie-Louise Aurumskjöld1, Kristina Ydström2, Anders Tingberg2

  • 1Medical Radiation Physics, Department of Translational Medicine, Lund University, Skåne University Hospital, Malmö, Sweden Marie-Louise.Aurumskjold@skane.se.

Acta Radiologica (Stockholm, Sweden : 1987)
|March 1, 2016
PubMed
Summary
This summary is machine-generated.

Iterative reconstruction methods significantly reduce computed tomography (CT) noise, improving image quality and enabling lower radiation doses for patients. Model-based iterative reconstruction (IMR) offers the greatest noise reduction compared to filtered back projection (FBP).

Keywords:
Computed tomography (CT)dose reductionimage qualityiterative reconstructionnoise

More Related Videos

Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging
04:54

Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging

Published on: June 16, 2023

4.0K
Hybrid µCT-FMT imaging and image analysis
13:45

Hybrid µCT-FMT imaging and image analysis

Published on: June 4, 2015

13.7K

Related Experiment Videos

Last Updated: Mar 25, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
06:33

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement

Published on: July 29, 2013

11.8K
Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging
04:54

Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging

Published on: June 16, 2023

4.0K
Hybrid µCT-FMT imaging and image analysis
13:45

Hybrid µCT-FMT imaging and image analysis

Published on: June 4, 2015

13.7K

Area of Science:

  • Radiology
  • Medical Imaging
  • Image Reconstruction

Background:

  • Increasing computed tomography (CT) use elevates patient radiation exposure.
  • Optimizing CT imaging is crucial for radiation dose reduction.
  • Iterative reconstruction (IR) methods enhance image quality and lower radiation dose.

Purpose of the Study:

  • Evaluate image quality based on reconstruction methods.
  • Assess patient dose reduction with hybrid and model-based iterative reconstruction.

Main Methods:

  • Utilized Catphan® 600 and anthropomorphic torso phantoms on a Philips Brilliance iCT.
  • Assessed image quality via CT numbers, noise, NPS, CNR, and resolution.
  • Reconstructed images using filtered back projection (FBP), hybrid (iDose⁴), and model-based iterative reconstruction (IMR).

Main Results:

  • iDose⁴ reduced noise by 15-45%; IMR reduced noise by 60-75% compared to FBP, independent of dose.
  • Noise Power Spectra (NPS) indicated altered noise distribution with different reconstruction methods.
  • Contrast-to-Noise Ratio (CNR) and low-contrast resolution improved with iDose⁴ and IMR.

Conclusions:

  • Hybrid and model-based iterative reconstruction methods offer significant potential for noise reduction and image quality improvement.
  • These advanced reconstruction techniques allow for lower radiation doses while maintaining diagnostic image quality.
  • Model-based iterative reconstruction (IMR) demonstrates superior noise reduction capabilities.