Data and methods to assess occupational exposure to personnel involved in cardiac catheterization procedures
Kostas Perisinakis1, Georgia Solomou2, John Stratakis2
1Department of Medical Physics, Medical School, University of Crete, P.O. Box 2208, Heraklion 71003, Crete, Greece; Department of Medical Physics, University Hospital of Heraklion, P.O. Box 1352, Heraklion 71110, Crete, Greece.
Purpose:
To provide normalized scatter exposure data and methods for reliable estimation of cumulative effective dose and eye-lens equivalent dose to personnel involved in fluoroscopically guided cardiac catheterization (FGCC) procedures.
Methods:
An anthropomorphic phantom was placed supine on the table of a modern digital C-arm angiographic system and 17 different fluoroscopic projections commonly employed during FGCC procedures were represented. Scatter exposure rates at the waist and eye level were measured for varying exposure parameters and position in the operating room. The effect of beam field size, patient size, use of radioprotective garments and small variations in projection angulation and table height on scatter radiation was investigated.
Results:
Apart from the position and use of radio-protective garments, radiation burden to operators during fluoroscopic guidance was found to remarkably depend beam field size (>45% reduction if a 10 × 10 cm(2) instead of 15 × 15 cm(2) fluoroscopy beam is used) and patient size (>25% increased scatter for obese patients). In contrast, the variation of measured scatter exposure from a given projection was found to be <10% when the source to skin distance was altered by ±10 cm or beam angulation of a specific projection was altered by ±10°.
Conclusion:
Presented scatter exposure data charts and methods allow for prospective and retrospective estimation of effective dose and eye-lens equivalent dose to personnel involved in any FGCC procedure. Projection specific maps of scatter exposure produced may enhance familiarization of involved medical staff to good radiation protection practice and optimization of working habits in the cardiac catheterization lab.
More Related Videos
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...