Optimizing the Exposure Indicator as a Dose Management Strategy in Computed Radiography
Purpose:
To investigate a technique for optimizing radiation dose and image quality for a computed radiography system.
Methods:
Entrance skin doses were measured for phantom models of the pelvis and lumbar spine imaged using the vendor's recommended exposure settings (ie, the reference doses) as well as doses above and below the vendor's recommended settings for both body parts. Images were assessed using visual grading analysis (VGA).
Results:
The phantom dosimetry results revealed strong positive linear relationships between dose and milliampere seconds (mAs), mAs and inverse exposure indicator (EI), and dose and inverse EI for both body parts. The VGA showed that optimized values of 16 mAs/EI = 136 for the anteroposterior (AP) pelvis and 32 mAs/EI = 139 for the AP lumbar spine did not compromise image quality.
Discussion:
Selecting optimized mAs reduced dose by 36% compared with the vendor's recommended mAs (dose) values.
Conclusion:
Optimizing the mAs and associated EIs can be an effective dose management strategy.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...