Magnetization transfer from inhomogeneously broadened lines (ihMT): Improved imaging strategy for spinal cord
Olivier M Girard1,2, Virginie Callot1,2, Valentin H Prevost1,2
1Aix-Marseille Université, CRMBM, UMR 7339 CNRS - AMU, Marseille, France.
Purpose:
Inhomogeneous magnetization transfer (ihMT) shows great promise for specific imaging of myelinated tissues. Whereas the ihMT technique has been previously applied in brain applications, the current report presents a strategy for cervical spinal cord (SC) imaging free of cerebrospinal fluid (CSF) pulsatility artifacts.
Methods:
A pulsed ihMT preparation was combined with a single-shot HASTE readout. Electrocardiogram (ECG) synchronization was used to acquire all images during the quiescent phase of SC motion. However ihMT signal quantification errors may occur when a variable recovery delay is introduced in the sequence as a consequence of variable cardiac cycle. A semiautomatic retrospective correction algorithm, based on repetition time (TR) -matching, is proposed to correct for signal variations of long T1 -components (e.g., CSF).
Results:
The proposed strategy combining ECG synchronization and retrospective data pairing led to clean SC images free of CSF artifacts. Lower variability of the ihMT metrics were obtained with the correction algorithm, and allowed for shorter TR to be used, hence improving signal-to-noise ratio efficiency.
Conclusion:
The proposed methodology enabled faster acquisitions, while offering robust ihMT quantification and exquisite SC image quality. This opens great perspectives for widening the in vivo characterization of SC physiopathology using MRI, such as studying white matter tracts microstructure or impairment in degenerative pathologies. Magn Reson Med 77:581-591, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
Magnetic Resonance Imaging
Spinal Cord: Cross-sectional Anatomy
Gray Matter and its Components
Central to the gray matter is...
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
The Spinal Cord