Use of Pipeline Flex is associated with reduced fluoroscopy time, procedure time, and technical failure compared with
Elizabeth J Le1, Timothy Miller2, Yafell Serulle2
1Department of Neurosurgery, University of Maryland Medical Center, Baltimore, Maryland, USA Department of Diagnostic Radiology, Neuroradiology, University of Maryland Medical Center, Baltimore, Maryland, USA.
Background:
Flow diversion with the Pipeline embolization device is a well-established method of intracranial aneurysm treatment. However, deployment of the first-generation device (Pipeline Classic) can be technically challenging. The Pipeline Flex contains the same flow-diverting stent with a modified delivery system.
Objective:
To compare procedural outcomes between the first-generation device (Pipeline Classic) and the Pipeline Flex.
Methods:
Thirty-eight of the first 40 consecutive patients who underwent intracranial aneurysm treatment with the Pipeline Flex and 58 of the most recent 60 consecutive patients who underwent treatment with the Pipeline Classic at our institution were evaluated. Patient demographics, aneurysm characteristics, technical procedural details, and early outcomes were analyzed.
Results:
The two groups were comparable for age, gender, and location of target aneurysms. Use of Pipeline Flex decreased procedure time by 44.2 min (p≤0.001) and fluoroscopy time by 22.0 min (p=0.001) compared with the Pipeline Classic. Similarly, radiation exposure was less in the Flex group with a mean difference of 3473.5 Gy cm2 (p=0.002), while contrast usage was decreased with a mean difference of 22.3 mL (p=0.007). These differences remained significant in multivariate regression analysis. Finally, the rate of device deployment failure was lower in the Flex group (7.1%) than in the Classic group (23.9%) (p=0.034).
Conclusions:
Use of Pipeline Flex significantly reduces the total procedure and fluoroscopy time, contrast usage, patient radiation exposure, and proportion of recaptured devices in comparison with the Pipeline Classic, probably owing to an enhanced delivery system that allows for more reliable and controlled deployment.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...