Aerosol Delivery Using Jet Nebulizer and Vibrating Mesh Nebulizer During High Frequency Oscillatory Ventilation: An
Tien-Pei Fang1,2, Hui-Ling Lin3, Shu-Hua Chiu1
11 Department of Respiratory Therapy, Chang Gung Memorial Hospital , Chiayi, Taiwan .
Background:
High frequency oscillatory ventilation (HFOV) is used in critically ill patients with severe hypoxemic respiratory failure. The purpose of this in vitro study was to determine the efficiency of aerosol delivery with different lung parameters during simulated neonatal, pediatric, and adult HFOV.
Methods:
Sensormedics 3100A/B ventilators were used to deliver infant, pediatric, and adult HFOV. Two types of aerosol generators were chosen for testing: 1) a continuous jet nebulizer (JN) with a unit-dose of 5.0 mg/2.5 mL salbutamol sulfate diluted into 4 mL, and 2) a vibrating mesh nebulizer (VMN) with salbutamol sulfate were run to completion of aerosol generation. Both aerosol devices were placed 1) between the ventilator circuit and the endotracheal tube (ETT) (proximal position); and 2) at the inlet of the heated humidifier (distal position) (n = 5). Drug was collected on a bacterial filter placed distal to the ETT, and the drug eluted and analyzed with a UV Spectrophotometer at 276 nm. T- test and ANOVA tests were used for comparison (p < 0.05).
Results:
The inhaled drug delivered by JN was 0%-0.6% of the nominal dose when placed at distal position, and 0%-3% at proximal position (p < 0.01), while the VMN was 0%-0.5% at distal and 8.6%-22.7% at proximal position (p < 0.01). Aerosol delivery during HFOV was greater with adult settings than pediatric and infant settings with VMN and JN (22.7%, 8.6%, and 17.4% respectively, p < 0.01). When the aerosol delivery device was placed at the distal position, negligible drug mass was observed (<0.5%), regardless of the nebulizer device used.
Conclusions:
During HFOV, aerosol delivery with the nebulizer placed at proximal was greater than placement distal from the ETT, with VMN delivering more drug than JN. The inhaled drug was delivery correlated positively with ETT size, MAP, and bias flow, and inversely proportional to power settings.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Inhaled Medications