[Inappropriate ICD shocks after collision with a pole].

Sing-Chien Yap1, Dominic A M J Theuns, Tamas Szili-Torok

  • 1Erasmus MC, afd. Cardiologie, Rotterdam.

Nederlands Tijdschrift Voor Geneeskunde
|March 17, 2016
PubMed
Summary

A patient received inappropriate implantable cardioverter-defibrillator (ICD) shocks due to lead fracture. Cine-fluoroscopy revealed the fracture, leading to lead extraction and system replacement.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Pulsed Field Ablation With a Variable Loop Circular Catheter in Atrial Fibrillation: Acute Outcomes From the VARIPURE Multicenter Study.

Heart rhythm·2026
Same author

Contemporary trends and outcomes in European ICD recipients: a 15-year analysis.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology·2026
Same author

Using monophasic potential mapping to visualize the Brugada substrate.

Heart rhythm·2026
Same author

Automated Cardiac Arrest Detection Using Wrist-Worn Photoplethysmography: External Validation in Patients With Induced Shockable Cardiac Arrest (DETECT-1b).

Circulation. Arrhythmia and electrophysiology·2026
Same author

Effectiveness of AF pulsed field ablation with a variable-loop circular catheter: 12-month VARIPURE results.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology·2026
Same author

The Cost Utility of Rhythm Control Treatment Sequences in Patients with Atrial Fibrillation: Anti-arrhythmic Drugs Versus Catheter Ablation.

PharmacoEconomics·2026

Area of Science:

  • Cardiology
  • Medical Device Technology
  • Diagnostic Imaging

Background:

  • Implantable cardioverter-defibrillators (ICDs) are crucial for managing life-threatening arrhythmias.
  • Lead integrity is essential for the proper functioning of ICD systems.
  • Lead fractures can lead to inappropriate shocks or failure to deliver therapy.

Observation:

  • A 60-year-old male experienced multiple inappropriate ICD shocks.
  • The patient's symptoms occurred after a physical collision.
  • Standard chest radiography did not clearly visualize the suspected lead issue.

Findings:

  • Cine-fluoroscopy identified a proximal lead fracture in the patient's ICD system.
  • The lead fracture was the cause of the inappropriate ICD shocks.
  • Surgical intervention was required to address the device malfunction.

Implications:

  • Cine-fluoroscopy is a valuable tool for diagnosing subtle lead fractures not apparent on radiography.
  • Prompt lead extraction and ICD system revision are necessary following lead fracture diagnosis.
  • This case highlights the importance of advanced imaging in troubleshooting ICD lead failures.