Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Normative NeuroFlexor data for detection of spasticity after stroke: a cross-sectional study
Gaia Valentina Pennati1, Jeanette Plantin2, Jörgen Borg2
1Division of Rehabilitation Medicine, Department of Clinical Sciences Karolinska Institutet, Danderyd University Hospital, Stockholm, SE-182 88, Sweden. gaia.pennati@ds.se.
Background And Objective:
The NeuroFlexor is a novel instrument for quantification of neural, viscous and elastic components of passive movement resistance. The aim of this study was to provide normative data and cut-off values from healthy subjects and to use these to explore signs of spasticity at the wrist and fingers in patients recovering from stroke.
Methods:
107 healthy subjects (age range 28-68 years; 51 % females) and 39 stroke patients (age range 33-69 years; 33 % females), 2-4 weeks after stroke, were assessed with the NeuroFlexor. Cut-off values based on mean + 3SD of the reference data were calculated. In patients, the modified Ashworth scale (MAS) was also applied.
Results:
In healthy subjects, neural component was 0.8 ± 0.9 N (mean ± SD), elastic component was 2.7 ± 1.1 N, viscous component was 0.3 ± 0.3 N and resting tension was 5.9 ± 1 N. Age only correlated with elastic component (r = -0.3, p = 0.01). Elasticity and resting tension were higher in males compared to females (p = 0.001) and both correlated positively with height (p = 0.01). Values above healthy population cut-off were observed in 16 patients (41 %) for neural component, in 2 (5 %) for elastic component and in 23 (59 %) for viscous component. Neural component above cut-off did not correspond well to MAS ratings. Ten patients with MAS = 0 had neural component values above cut-off and five patients with MAS ≥ 1 had neural component within normal range.
Conclusion:
This study provides NeuroFlexor cut-off values that are useful for detection of spasticity in the early phase after stroke.