Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study
Sebastian Winklhofer1,2, Jack W Lambert1, Zhen Jane Wang1
1Department of Radiology and Biomedical Imaging, University of California San Francisco, 505 Parnassus Ave, M-372, Box 0628, San Francisco, CA, 94143-0628, USA.
Objective:
The purpose of the study was to assess the ability of rapid-kV switching (rs) dual-energy computed tomography (DECT) to reduce peristalsis-related streak artifact.
Methods:
rsDECT images of 100 consecutive patients (48 male, 52 female, mean age 57 years) were retrospectively evaluated in this institutional review board-approved study. Image reconstructions included virtual monochromatic 70 and 120 keV images, as well as iodine(-water) and water(-iodine) material decomposition images. We recorded the presence and severity of artifacts qualitatively (4-point scale) and quantitatively [iodine/water concentrations, Hounsfield units, gray scale values (GY)] and compared to corresponding unaffected reference tissue. Similar measures were obtained in DECT images of a peristalsis phantom. Wilcoxon signed-rank and paired t tests were used to compare results between different image reconstructions.
Results:
Peristalsis-related streak artifacts were found in 49 (49%) of the DECT examinations. Artifacts were significantly more severe in 70, 120, and water(-iodine) images than in iodine(-water) images (qualitative readout P < 0.001, each). Quantitative measurements were significantly different between the artifact and the reference tissue in 70, 120 keV, and water(-iodine) images (P < 0.001 for both HU and GY for each image reconstruction), but not significantly different in iodine(-water) images (iodine concentrations P = 0.088 and GY P = 0.111). Similar results were seen in the peristalsis DECT phantom study.
Conclusions:
Peristalsis-related streak artifacts seen in 70, 120 keV, and water(-iodine) images are substantially reduced in iodine(-water) images at rsDECT.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography