Virtual monoenergetic imaging for metal artifact reduction in dental implant surgery using photon-counting detector
Adib Al-Haj Husain1,2,3,4, Victor Mergen5, Thomas Sartoretti5
1Department of Cranio-Maxillofacial and Oral Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Purpose:
This ex vivo study was performed to determine the optimal energy level for virtual monoenergetic images (VMIs) generated with photon-counting detector computed tomography (PCD-CT) to minimize metal artifacts from dental implants.
Materials And Methods:
Twelve implants from various manufacturers were placed in 6 pig mandibles and scanned with PCD-CT. VMIs were reconstructed at energy levels from 70 keV to 150 keV in 20-keV increments. Three readers with varying experience qualitatively assessed the image quality, artifact burden, and diagnostic interpretability of peri-implant soft and hard tissues using a 5-point discrete visual scale. Objective analyses included quantitative line profile analysis of implant-induced artifacts. Descriptive statistics were calculated, and inter-reader agreement was assessed using percentage agreement and the Krippendorff alpha coefficient.
Results:
Qualitative analysis demonstrated excellent image quality for VMIs at ≥110 keV (median=5), with minimal artifacts observed at 130-150 keV. In contrast, lower-energy VMIs (70-90 keV) showed inferior performance due to artifact-related limitations in diagnostic interpretability. Inter-reader agreement ranged from moderate to perfect, with perfect reliability (α=1) for VMIs ≥110 keV. Quantitative line-profile analysis confirmed reduced artifact burden at higher energy levels, particularly for VMIs ≥110 keV.
Conclusion:
VMI at energy levels ≥110 keV on PCD-CT reduced dental implant-related metal artifacts and offered excellent image quality, including assessment of both peri-implant soft and hard tissues. These findings suggest that optimized PCD-CT VMI may enhance postoperative follow-up imaging. Future in vivo studies are warranted to validate these findings in clinical practice.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
X-ray Imaging