Related Experiment Video
Updated: Mar 24, 2026

Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
Phase correlation applied to the 3D registration of CT and CBCT image volumes
Daniel Foley1, Daniel J O'Brien1, Luis León-Vintró2
1St. Luke's Radiation Oncology Network, Highfield Road, Rathgar, Dublin 6, Ireland; School of Physics, University College Dublin, Belfield, Dublin 4, Ireland.
Purpose:
In this study, a 3D phase correlation algorithm was investigated to test feasibility for use in determining the anatomical changes that occur throughout a patient's radiotherapy treatment. The algorithm determines the transformations between two image volumes through analysis in the Fourier domain and has not previously been used in radiotherapy for 3D registration of CT and CBCT volumes.
Methods:
Various known transformations were applied to a patient's prostate CT image volume to create 12 different test cases. The mean absolute error and standard deviation were determined by evaluating the difference between the known contours and those calculated from the registration process on a point-by-point basis. Similar evaluations were performed on images with increasing levels of noise added. The improvement in structure overlap offered by the algorithm in registering clinical CBCT to CT images was evaluated using the Dice Similarity Coefficient (DSC).
Results:
A mean error of 2.35 (σ = 1.54) mm was calculated for the 12 deformations applied. When increasing levels of noise were introduced to the images, the mean errors were observed to rise up to a maximum increase of 1.77 mm. For CBCT to CT registration, maximum improvements in the DSC of 0.09 and 0.46 were observed for the bladder and rectum, respectively.
Conclusions:
The Fourier-based 3D phase correlation registration algorithm investigated displayed promising results in CT to CT and CT to CBCT registration, offers potential in terms of efficiency and robustness to noise, and is suitable for use in radiotherapy for monitoring patient anatomy throughout treatment.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...