Performance of a Blood Glucose Monitoring System in a Point-of-Care Setting
Cornelia Ottiger1, Nicole Gygli1, Andreas R Huber1
1Cantonal Hospital Aarau AG, Aarau, Switzerland.
Journal of Diabetes Science and Technology
|March 19, 2016
Summary
The CONTOUR® XT-BGMS (CXT-BGMS) meets quality control standards for point-of-care testing (POCT) in hospitals. This blood glucose monitoring system (BGMS) demonstrates high accuracy and compliance with international guidelines.
Area of Science:
- Clinical Chemistry
- Medical Device Technology
- Point-of-Care Testing (POCT)
Background:
- Point-of-care testing (POCT) requires stringent quality control for reliable patient results.
- Hospital settings demand accurate and compliant blood glucose monitoring systems (BGMS).
- International guidelines, including ISO15197:2013, RiliBÄK, and QUALAB, set critical accuracy and quality benchmarks for BGMS.
Purpose of the Study:
- To evaluate the CONTOUR® XT-BGMS (CXT-BGMS) for accuracy and compliance in a hospital POCT setting.
- To assess CXT-BGMS adherence to German (RiliBÄK) and Swiss (QUALAB) quality control guidelines.
- To verify CXT-BGMS performance against ISO15197:2013 accuracy criteria.
Main Methods:
- A single-center study conducted in Switzerland.
- Testing of 105 venous blood samples from hospitalized patients.
- Comparison of CXT-BGMS results against the hexokinase reference method.
- Daily quality control measurements using control solutions to assess POCT guideline compliance.
Main Results:
- CXT-BGMS achieved 98.41% accuracy, meeting ISO15197:2013 criteria.
- All control measurements complied with RiliBÄK guidelines (±11% of target values, within RMSE limits).
- All control measurements complied with QUALAB guidelines (±10% of target values).
Conclusions:
- The CONTOUR® XT-BGMS (CXT-BGMS) demonstrates compliance with German (RiliBÄK) and Swiss (QUALAB) quality control guidelines.
- CXT-BGMS meets the ISO15197:2013 accuracy requirements for blood glucose monitoring systems in hospital POCT settings.
- The study confirms the reliability and accuracy of CXT-BGMS for routine use in clinical practice.
Related Concept Videos
Errors occurring during blood pressure monitoring
1.7K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.7K
Pulse rhythm
1.6K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.6K
Assessing Blood pressure in the Leg
14.4K
Proper measurement of leg blood pressure is a critical skill for healthcare providers, ensuring precise and reliable readings. When performed correctly, this procedure informs patient care and enhances the efficacy of interventions. The following text outlines step-by-step guidelines to measure blood pressure in the leg, providing clarity and ease of understanding for practitioners.
Preparation:
Preparation:
14.4K
SBAR II: Application of SBAR
6.8K
SBAR is an effective communication tool used by healthcare professionals to communicate patient information accurately. SBAR stands for Situation, Background, Assessment, and Recommendation. For a better understanding, an example is given below.
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
6.8K
Hypoglycemia and Glucagon
1.3K
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
1.3K
Assessment of blood pressure in brachial artery(one-step method)
1.4K
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
1.4K