Related Experiment Video
Updated: Mar 24, 2026

The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
Accurate nonfluoroscopic guidance and tip location of peripherally inserted central catheters using a conductance
Mark C Svendsen1, David Birrer1, Benjamin Jansen1
1Department of Biomedical Engineering, Indiana University-Purdue University Indianapolis, Indianapolis, Ind.
Background:
Bedside placement of peripherally inserted central catheters (PICCs) may result in navigation to undesirable locations, such as the contralateral innominate or jugular vein, instead of the superior vena cava or right atrium. Although some guidance and tip location tools exist, they have inherent limitations because of reliance on physiological measures (eg, chest landmarks, electrocardiogram, etc), instead of anatomical assessment (ie, geometric changes in the vasculature). In this study, an accurate, anatomically based, non-X-ray guidance tool placed on a novel 0.035" conductance guidewire (CGW) is validated for PICC navigation and tip location.
Methods:
The CGW system uses electrical conductance recordings to assess changes in vessel cross-sectional area to guide navigation of the PICC tip. Conductance rises and oscillates when going in the correct direction to the superior vena cava/right atrium, but drops when going in the incorrect direction away from the heart. Bench and in vivo studies in six swine were used to confirm the accuracy and repeatability of the PICC placement at various anatomical locations. The PICC tip location was confirmed by direct visualization vs the desired location.
Results:
CGW PICC guidance was highly accurate and repeatable with virtually no difference between actual and desired catheter tip location. The difference between the CGW PICC location vs the desired target was -0.07 ± 0.07 cm (6.6% error) on the bench and 0.04 ± 0.10 cm (5% error) in vivo. No complications or adverse events occurred during CGW usage.
Conclusions:
The CGW provides an anatomically based, reproducible, and clinically significant method for PICC navigation and tip location that can improve accuracy, decrease the wait time prior to therapy delivery, decrease cost, and minimize the need for X-ray. These findings warrant clinical evaluation of this navigation tool for PICC line placement.
More Related Videos
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management