[Development of a Lead-covered Case for a Wireless X-ray Output Analyzer to Perform CT Half-value Layer Measurements]
Hirokazu Akaishi1, Hiromitsu Takeda, Yoshiyuki Kanazawa
1Biomedical Research, Education and Instrumentation Center, Sapporo Medical University.
Abstract:
Measurement of the half-value layer (HVL) is a difficult task in computed tomography (CT) , because a nonrotating X-ray tube must be used. The purpose of this study is to develop a lead-covered case, which enables HVL measurements with a rotating CT X-ray tube. The lead-covered case was manufactured from acrylic and lead plates, which are 3 mm thick and have a slit. The slit-detector distance can be selected between 14 mm and 122 mm. HVL measurements were performed using a wireless X-ray output analyzer "Piranha." We used the following exposure conditions: tube voltages of 80, 100, and 120 kV; a tube current of 550 mA; and an exposure time of 1.0 s. The HVLs were measured by using the following two methods: (a) Nonrotating method-a conventional method that uses the nonrotating exposure mode. (b) Rotating method-a new method that uses the lead-covered case and the rotating exposure mode. As a result, when the slit-detector distance was 58 mm, the HVL values obtained by the nonrotating and rotating methods were 4.38 and 4.24 mmAl at 80 kV, 5.51 and 5.37 mmAl at 100 kV, 6.61 and 6.48 mmAl at 120 kV, respectively. A lead-covered case, which enables the measurement of the HVL in a rotating X-ray tube, was developed. The case is useful in measuring the HVLs at facilities that cannot fix the X-ray tube.