Strategies for calibrating a subcutaneous glucose sensor.

G Velho1, P Froguel, D R Thevenot

  • 1Service de Diabétologie, Hôtel-Dieu, Paris, France.

Biomedica Biochimica Acta
|January 1, 1989
PubMed
Summary

Accurate glucose sensor calibration is vital for subcutaneous implants. In vivo calibration, unlike in vitro, accurately estimates subcutaneous glucose levels by reflecting blood glucose concentrations in rats.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Relationship between renal capacity to reabsorb glucose and renal status in patients with diabetes.

Diabetes & metabolism·2020
Same author

Dairy consumption is associated with lower plasma dihydroceramides in women from the D.E.S.I.R. cohort.

Diabetes & metabolism·2019
Same author

Temporality in chronic diseases and adherence to long-term therapies: From philosophy to science and back.

Diabetes & metabolism·2018
Same author

Holistic psychosocial determinants of adherence to medication in people with type 2 diabetes.

Diabetes & metabolism·2018
Same author

Migrant adults with diabetes in France: Influence of family migration.

Journal of clinical & translational endocrinology·2017
Same author

Clinical inertia and its impact on treatment intensification in people with type 2 diabetes mellitus.

Diabetes & metabolism·2017

Area of Science:

  • Biomedical Engineering
  • Biosensors
  • Medical Devices

Background:

  • Subcutaneous glucose sensor calibration is critical for accurate glucose concentration estimation.
  • In vitro calibration methods may not accurately reflect in vivo conditions for subcutaneous glucose sensors.

Purpose of the Study:

  • To compare subcutaneous glucose concentration estimations using in vitro versus in vivo calibration parameters.
  • To determine the optimal calibration method for subcutaneous glucose sensors in rats.

Main Methods:

  • In vitro calibration performed in phosphate buffer at 37°C.
  • In vivo calibration determined by correlating sensor output changes with blood glucose changes induced by insulin and glucose infusions.
  • Comparison of subcutaneous glucose estimations using both calibration methods.

Main Results:

  • In vitro calibration resulted in significantly lower apparent subcutaneous glucose concentrations compared to blood glucose levels.
  • In vivo calibration accurately reflected blood glucose levels, with a five-minute time lag, across hypoglycemia and hyperglycemia ranges.
  • Two-point in vivo calibration demonstrated superior performance in estimating subcutaneous glucose.

Conclusions:

  • Two-point in vivo calibration is essential for accurate subcutaneous glucose concentration estimation.
  • In vitro calibration parameters are insufficient for reliable subcutaneous glucose monitoring.
  • In vivo calibration improves the accuracy of implanted glucose sensor readings.