A modified device for continuous non-invasive blood pressure measurements in humans under hyperbaric and/or
René van der Bel1, Bart C Sliggers2, Marc J van Houwelingen3
1Department of Internal Medicine, Academic Medical Center, University of Amsterdam, The Netherlands.
Background:
It would be desirable to safely and continuously measure blood pressure noninvasively under hyperbaric and/or hyperoxic conditions, in order to explore haemodynamic responses in humans under these conditions.
Methods:
A systematic analysis according to 'failure mode and effects analysis' principles of a commercially available beat-by-beat non-invasive blood pressure monitoring device was performed using specifications provided by the manufacturer. Possible failure modes related to pressure resistance and fire hazard in hyperbaric and oxygen-enriched environments were identified and the device modified accordingly to mitigate these risks. The modified device was compared to an unaltered device in five healthy volunteers under normobaric conditions. Measurements were then performed under hyperbaric conditions (243 kPa) in five healthy subjects.
Results:
Modifications required included: 1) replacement of the carbon brush motorized pump by pressurized air connected through a balanced pressure valve; 2) modification of the 12V power supply connection in the multiplace hyperbaric chamber, and 3) replacement of gas-filled electrolytic capacitors by solid equivalents. There was concurrence between measurements under normobaric conditions, with no significant differences in blood pressure. Measurements under pressure were achieved without problems and matched intermittent measurement of brachial arterial pressure.
Conclusion:
The modified system provides safe, stable, continuous non-invasive blood pressure trends under both normobaric and hyperbaric conditions.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: