Measurement of Temperature Dependent Apparent Specific Heat Capacity in Electrosurgery
Wafaa Karaki1, Ali Akyildiz1, Diana-Andra Borca Tasciuc1
1Department of Mechanical, Aerospace and Nuclear Engineering, Rensselaer Polytechnic University, Troy, NY.
Abstract:
This paper reports on the measurement of temperature dependent apparent specific heat of ex-vivo porcine liver tissue during radiofrequency alternating current heating for a large temperature range. The difference between spatial and temporal evolution of experimental temperature, obtained during electrosurgical heating by infrared thermometry, and predictions based on finite element modeling was minimized to obtain the apparent specific heat. The model was based on transient heat transfer with internal heat generation considering heat storage along with conduction. Such measurements are important to develop computational models for real time simulation of electrosurgical procedures.
More Related Videos
08:50Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
10:23Author Spotlight: Computing the Effects of a Local Radiofrequency Hyperthermia Intervention on Tumor Biomechanics
Published on: December 1, 2023
Related Concept Videos
Specific Heat
For example, increasing the temperature of one gram of water by 1°C requires one calorie of heat energy and can be written as 1 cal/g-°C, or...
Quantifying Heat
Heat Capacities of an Ideal Gas I
Molar heat capacity quantifies the ratio of the amount of heat added (or removed) to increase (or decrease) the...
Calorimetry