Collection, Expansion, and Differentiation of Primary Human Nasal Epithelial Cell Models for Quantification of Cilia Beat Frequency
Published on: November 10, 2021
5.0K
A Low-Cost Method of Ciliary Beat Frequency Measurement Using iPhone and MATLAB: Rabbit Study
Jason J Chen1, Bryan T Lemieux1, Brian J F Wong2
1Beckman Laser Institute, University of California-Irvine, Irvine, California, USA.
Summary
This study developed a cost-effective method to measure ciliary beat frequency (CBF) using cellphone videos and MATLAB. The results show this technique accurately reflects ciliary function, offering a fast and affordable alternative for research.
Area of Science:
- Biomedical Engineering
- Otolaryngology Research
- Respiratory Physiology
Background:
- Accurate measurement of ciliary beat frequency (CBF) is crucial for understanding mucociliary clearance.
- Existing methods for CBF measurement can be expensive and time-consuming.
- Consumer-grade technology offers potential for more accessible research tools.
Purpose of the Study:
- To establish a method for determining ciliary beat frequency (CBF) utilizing a standard cellphone camera and MATLAB software.
- To assess the accuracy and efficacy of this novel, low-cost approach compared to traditional techniques.
Main Methods:
- Ex vivo rabbit tracheal samples were analyzed under phase-contrast microscopy.
- Cilia movement was recorded using an iPhone 6 at 240 frames per second (fps).
- MATLAB was employed for video analysis, including Fourier analysis at various frame rates (240, 120, 60, 30 fps) and manual verification.
Main Results:
- Ciliary beat frequency (CBF) was significantly higher at 37°C (9.08 ± 0.49 Hz) compared to 23°C (5.03 ± 0.4 Hz).
- Formalin-fixed samples showed no measurable ciliary activity.
- No statistically significant differences in CBF measurements were found across tested frame rates (240, 120, 60, 30 fps) or compared to manual counting at 23°C.
Conclusions:
- The developed cellphone-based method provides accurate ciliary beat frequency (CBF) measurements, comparable to visual inspection.
- All tested acquisition frame rates proved effective.
- This technique presents a rapid, inexpensive, and viable alternative to conventional CBF measurement protocols.
Keywords:
CBFHSVAMATLABSAVASisson-Ammons video analysisciliary beat frequencyhigh-speed video analysisiPhonephone camera