Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Can higher end tonometers be used interchangeably in routine clinical practice?
Kunjan Jayantilal Patel1, Saurabh P Jain, Priti R Kapadia
1Department of Ophthalmology, Government Medical College, Surat, Gujarat, India.
Intraocular pressure (IOP) measurement agreement varied between tonometers, with clinically unacceptable limits. For consistent glaucoma patient monitoring, use a single tonometer at each follow-up.
Area of Science:
- Ophthalmology
- Glaucoma Management
- Ophthalmic Instrumentation
Background:
- Accurate intraocular pressure (IOP) measurement is critical for glaucoma management.
- Several instruments exist for IOP assessment, raising questions about their interchangeability.
- Understanding the agreement between different tonometers is essential for clinical practice.
Purpose of the Study:
- To evaluate the agreement in intraocular pressure (IOP) measurements among noncontact tonometer (NCT), rebound tonometer (RBT), Goldmann applanation tonometer (GAT), and dynamic contour tonometer (DCT).
- To determine if IOP measurements from newer tonometers are interchangeable with the Goldmann applanation tonometer (GAT), considered the gold standard.
Main Methods:
- A study involving 499 eyes from 250 patients over 24 months.
- Intraocular pressure (IOP) was measured sequentially using NCT, RBT, GAT, and DCT.
- Bland-Altman plots were utilized to assess agreement, with GAT serving as the reference standard.
Main Results:
- The mean IOP readings across all devices (NCT, RBT, GAT, DCT) were similar, approximately 15.9-16.0 mm Hg.
- Limits of agreement between GAT and DCT, NCT, and RBT were found to be ±5.2 to ±5.4 mm Hg.
- While a strong correlation existed, the calculated limits of agreement were deemed clinically unacceptable.
Conclusions:
- Newer tonometers show a strong correlation with the Goldmann applanation tonometer (GAT).
- However, the limits of agreement between these devices are clinically unacceptable for interchangeable use.
- Consistent use of a single tonometer per patient during follow-up visits is recommended in glaucoma clinics.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Uncertainty in Measurement: Reading Instruments