Related Experiment Video
Updated: Jul 19, 2026

Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
Multiple Site Fracture of Both Rods in a Malleable Penile Implant
Marcelo Almeida Pinheiro1, Haroldo Brasil Barroso Filho1, Francisco José Cabral Mesquita2
1School of Medicine, Unichristus University, 60190-060 Fortaleza, CE, Brazil.
Abstract:
Penile prosthesis implant is the definitive treatment for refractory erectile dysfunction. Fracture of malleable prosthesis is rarely described due to its low incidence. We describe a case of multiple, bilateral fracture of a malleable penile implant, ten years after implantation. After the diagnosis, a review surgery was performed and the implants were replaced. No corporal rupture or urethral lesion was observed. Review of the literature shows few articles reporting penile implant fractures, and to our knowledge no other article has described multiple, bilateral fractures of a penile prosthesis.