An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Biomechanical Evaluation of 6.5-mm Cannulated Screws
Benjamin C Taylor1, Alan S Litsky, Kevin J Pugh
1Department of Orthopaedic Surgery, Grant Medical Center, Columbus, Ohio. drbentaylor@gmail.com.
Abstract:
Although biomechanical and clinical evidence exists regarding smaller compression screws, biomechanical data regarding the larger headless screws are not currently available. Headed and headless 6.5-mm cannulated compression screws were examined, with analysis of interfragmentary compression, insertion torque, and resistance of the construct to a shear force. No significant differences were seen between the maximum insertion torque of the headless or headed screws. Maximum and steady-state compression forces were also not significantly different between groups. Countersinking the headless model 2 mm led to a 77.01% decrease in steady-state compression levels. Shear testing did not reveal any significant differences in peak load at ultimate failure, specimen stiffness, or final block displacement, although a trend to increased peak load and stiffness was seen with the headless specimens.