Background And Study Aims:
Endoscopic interventions require accurate and precise control of the endoscope tip. The endoscope tip response depends on a cable pulling system, which is known to deliver a significantly nonlinear response that eventually reduces control. It is unknown whether the current technique of endoscope tip control is adequate for a future of high precision procedures, steerable accessories, and add-on robotics. The aim of this study was to determine the status of the tip response of endoscopes used in clinical practice.
Materials And Methods:
We evaluated 20 flexible colonoscopes and five gastroscopes, used in the endoscopy departments of a Dutch university hospital and two Dutch teaching hospitals, in a bench top setup. First, maximal tip bending was determined manually. Next, the endoscope navigation wheels were rotated individually in a motor setup. Tip angulation was recorded with a USB camera. Cable slackness was derived from the resulting hysteresis plot.
Results:
Only two of the 20 colonoscopes (10 %) and none of the five gastroscopes reached the maximal tip angulation specified by the manufacturer. Four colonoscopes (20 %) and none of the gastroscopes demonstrated the recommended cable tension. Eight colonoscopes (40 %) had undergone a maintenance check 1 month before the measurements were made. The tip responses of these eight colonoscopies did not differ significantly from the tip responses of the other colonoscopes.
Conclusion:
This study suggests that the majority of clinically used endoscopes are not optimally tuned to reach maximal bending angles and demonstrate adequate tip responses. We suggest a brief check before procedures to predict difficulties with bending angles and tip responses.
More Related Videos
Related Concept Videos
Bending of Material: Problem Solving
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Bending of Curved Members - Strain Analysis
The important part of bending analysis for such a member...