Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Moist Heat Disinfection and Revisiting the A0 Concept
Abstract:
Moist heat is employed in the medical device, pharmaceutical, and food processing industries to render products and goods safe for use and human consumption. Applications include its use to pasteurize a broad range of foods and beverages, the control of microbial contamination of blood products, and treatment of bone tissue transplants and vaccines. In the pharmaceutical industry, water heated to 65°C to 80°C is used to sanitize high-purity water systems. In healthcare, it has been employed for decades to disinfect patient care items ranging from bedpans to anesthesia equipment. There is a good understanding of the conditions necessary to achieve disinfection of microorganisms at temperatures ranging from 65°C to 100°C. Based on this information, the efficacy of moist heat processes at a range of exposure times and temperatures can be quantified based on mathematical models such as the A0 calculation. While the A0 concept is recognized within the European healthcare community, it has yet to be widely adopted within the United States. This article provides information regarding the A0 concept, a brief overview of the classification of thermal disinfection for use with healthcare applications within the United States, and recent data on reinvestigating the thermal disinfection of a selected panel of microorganisms and a mixed culture biofilm.
More Related Videos
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
14:17The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Temperature
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Aqueous Solutions and Heats of Hydration
When ionic compounds dissolve in water, the ions in the solid separate and disperse uniformly throughout the solution because water molecules surround and solvate the ions, reducing the strong electrostatic forces between them. This process...