Related Experiment Video
Updated: Mar 22, 2026

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Heater-cooler units: contamination of crucial devices in cardiothoracic surgery
T Götting1, S Klassen1, D Jonas1
1Infection Prevention and Hospital Epidemiology, University Medical Center Freiburg, Freiburg, Germany.
Background:
Several cases of Mycobacterium chimaera infection have recently been reported in cardiosurgical patients. So-called heater-cooler units (HCUs) used in cardiosurgical procedures are suspected to be the reservoir for pathogen growth and dissemination.
Aim:
To assess the contamination status of HCUs at our facility.
Methods:
Air sampling for mycobacteria was conducted at different distances from the machines and in the area around the operating table. Air sampling was also conducted for non-fermenters as a surrogate parameter for water-associated pathogens.
Findings:
Mycobacterium chimaera was detected in the water tanks of the HCUs. When the devices were operating, M. chimaera was also found in their exhaust air, as well as in the area around the operating table. Non-fermenters were identified at different distances from the running HCU and the area around the operating table. Cultures remained negative when the devices were switched off.
Conclusions:
Exhaust air from HCUs may be a pathway of pathogen transmission to patients undergoing open chest heart surgery. Although, for technical reasons, relocation of HCUs is difficult to achieve, only strict separation of the HCU from the operating room appears to enhance patient safety. Using non-fermenters as a surrogate parameter may be considered a viable option for a timely risk assessment. The design of HCUs should be modified to keep susceptibility to contamination at a minimum.
More Related Videos
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Tracheostomy Suctioning II: Procedure